Tissue Response:Nerve: Page 4

Photograph showing early versions of a bipolar Huntington helix electrode. Subsequent designs use fewer wraps and a straight section, rather than wraps, between the stimulating contacts. Stimulating contacts can be seen between the ends of the first and second wraps, counting from the left hand side, and the sixth and seventh wraps.

A helical electrode implanted on a nerve is shown in this figure. Long arrow points to an angulation in the nerve at one end of the electrode, at the lead junction.
Kinks like this can cause mechanical damage to the nerve.

 